Glove Box

From NanoWiki

(Difference between revisions)
Jump to: navigation, search
Line 1: Line 1:
[[Category:Manuals]]
[[Category:Manuals]]
-
'''Please read these instructions before using the Glove Box, and log every operation in the [[Glove Box history]]'''
+
[[File:Cow_small.jpg|right]]
-
'''All users require training from a trained user (Feri, Fábián Gábor, Szirmai Péter). Use is allowed only if the trainer approves to that. Any violation results in being banned from use.'''
+
'''Please read these instructions before using the Glove Box, and log every operation in the Glove box logbook!'''
-
[[File:Cow_small.jpg|right]]
+
'''All users require training from a trained user (Feri, Fábián Gábor, Szirmai Péter). Use is allowed only if the trainer approves to that. Any violation results in being banned from use.'''
==Normal operation==
==Normal operation==
-
[[File:Glovebox.jpg|left|800px|Left: Good (GB1 and VPG open, RKM active) Right: Not good]]
+
[[File:Glovebox.jpg|center|800px|thumb|Left: Good (GB1 and VPG open, RKM active) Right: Not good]]
===Checking before the work, also check daily when box is unused===
===Checking before the work, also check daily when box is unused===

Revision as of 12:21, 31 August 2011

Cow small.jpg

Please read these instructions before using the Glove Box, and log every operation in the Glove box logbook!

All users require training from a trained user (Feri, Fábián Gábor, Szirmai Péter). Use is allowed only if the trainer approves to that. Any violation results in being banned from use.

Contents

Normal operation

Left: Good (GB1 and VPG open, RKM active) Right: Not good

Checking before the work, also check daily when box is unused

  • Circulator and water cooling is running. Never press Regeneration button on the panel.
  • Box pressure is 3-8 mbar, O2, H2O levels are below 0.1 ppm. If any of above is not like that, do not work!!! Report to Feri, 30-452-4705
  • Check pressure in Ar bottle-station. If below 10 bar, change bottle. PURGE!!!
  • Turn on the vacuum pump before use, wait until end vacuum goes to below < 0.1 mbar (indicator is on the Leybold station). The vacuum pump should not run when the box is not in use. Switch on/off with a separate switch on the floor near the pump.

Working

  • check if inner door of both antechambers are closed
  • fill up the used chamber from box (keep a slight 50-100 mbar pressure under the box, this prevents that residual air in the otherwise evacuated antechamber goes into the box)
  • open chamber door. NEVER TOUCH the inside with pure hands, use gloves.
  • place sample close to inner end. Never put in any solvents. Samples in containers should be a bit opened!
  • BBe careful with the black valve handle, experience shows that if not gently handled it will leak!
  • Evacuate to cc. 0.1 mbar (2 min. for small, 10 min for large chamber). Refill/repump twice to 500 mbar (max to 950 mbar, this prevents that residual air in the otherwise evacuated antechamber goes into the box). (Total: 6 min for small, 30 min for large chamber).
  • Use gloves with talc. Use a shirt with a LONG SLEEVE. Use talc on naked arms. Remove rings, watches, bracelets, piercings etc.
  • Go inside slowly so that box can react the pressure change. Same for coming out.

Leave box after work:

  • Always close inner doors. Careful, only the handle part of the small chamber door should be touched. Glove can be hurt!!!
  • Switch off light
  • Both Chambers are closed from in-outside
  • Chambers are pumped out to vacuum (< 0.1 mbar) and closed from pump, i.e. both chamber valves are in the close position.
  • Check O2, H2O levels. If not below 0.5 ppm, note the time it takes to go down. Report to a trained user!
  • Switch off the vacuum pump if you do not use it for a day at least.

Regeneration (if O2 or H2O level does not go below 0.1 ppm after 1 day without use)

  • Secure all important samples into chambers, pump the chambers out, close.
  • Switch off Circulator
  • set 0.3 bar on Ar-H2
  • turn on Regeneration, process cannot be stopped!
  • 1/2 bottle (100 bar) of Ar-H2 per regeneration is needed
  • Status can be monitored by clicking on the absorbent and checking "Status"

Regular service:

  • H2O sensor clean: 2-3 months
  • Gloves: every second year or when worn out. Small leaks around mounting ring. Change glove O-rings as well.
  • Particle filters: once per year
  • Small-large antechamber O-rings: when looks worn out (after cc. 3 years)